Expression Information Portal - IP5 (secondary display control unit for Expression MRI patient monitoring system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01029-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-09-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Continuous use of the ip5 for more than 24 days, without power cycling, results in a graphical user interface (display screen) error, which causes the numeric data on the display to freeze and disables some menu keys. waveform data displayed and audible alarms are not impacted by this problem and continue to function properly. this could result in delay in treatment of the patient caused by the displayed numeric value not correlating with the patient's status.This is an update to the previous recall for product correction (tga ref.: rc-2014-rn-00349-1) which informed end users to ensure that the ip5 should be power cycled every 24 hours to prevent the above issue from occurring. now a permanent software solution is available.
  • 조치
    A Philips representative will contact the customers and all affected devices will have a software update installed in order to address the problem. After the software update is installed, if the device is allowed to run for an extended period of time, the device will begin to notify users of the need to power cycle the device using various degrees of warning “restart required” prompts (dialog boxes). If the device is not power cycled when this time expires, the device will power down and sound a tone to inform users of the action. Until the software solution is implemented, users should ensure that the IP5 is power cycled once every day. This action has been closed-out on 5/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Expression Information Portal - IP5 (secondary display control unit for Expression MRI patient monitoring system)Product Code: 865471 Service Number 453564251511Serial Numbers: US23200375, US23200376, US23200377ARTG Number: 189095
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA