EZee-Nest Insert Cups used with ADVIA Centaur CP. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00389-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified an issue with use of the ezee-nest insert cups due to 2 inaccuracies that exist in the advia centaur cp operator’s guide, appendix d, pre-set tube types. the 1st inaccuracy refers to the sample volume supported by the system when using the ezee-nest cup to process samples. the operator's guide states that sample volumes greater than 50ul can be used; however volumes greater than 150ul cannot be used when using an ezee-nest insert cup. the 2nd inaccuracy refers to the use of the universal cup adaptor with an ezee-nest insert cup in 65-mm sample racks. siemens has determined that the height of an ezee-nest insert cup with universal cup adaptor fitted on a 65-mm sample rack will not fit into the sample compartment. there is no impact on patient samples as a result of these issues. there is the potential for loss of aspirated sample (up to 250 ìl including the dead volume in the insert cup) & a delay in providing patient test results due to this issue.
  • 조치
    Siemens is advising users to not use EZee-Nest Insert Cups with assays that require sample volumes greater than 150µL or the Universal Cup Adapter with the EZnee-Nest insert cups in 65mm sample racks. Users should instead use the Siemens Sample cup along with the required 100mm sample rack and 100mm Siemens cup Tube Setting for assays that require sample volumes larger than 150µL.

Device

  • 모델명 / 제조번호(시리얼번호)
    EZee-Nest Insert Cups used with ADVIA Centaur CP. An in vitro diagnostic medical device (IVD)Siemens Material Number (SMN): 10309710ARTG Number: 175890
  • Manufacturer

Manufacturer