Factor IX coagulation studies on the BCS / BCS XP Automated Blood Coagulation Analyser (Actin, Actin FSL and Pathromtin SL) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00368-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-04-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has detected a situation with bcs / bcs xp analysers regarding factor ix assay determinations for haemophilia b patients. internal investigations showed discrepancies for siemens bcs / bcs xp within dilution studies. siemens has confirmed that samples at the lower measurement range were found elevated on the bcs/bcs xp system. patient results in the lower measuring range may be misclassified within a higher factor ix level. patients that have been evaluated for hereditary factor ix deficiency disease for the first time may have to be re-evaluated.A falsely elevated factor ix level may lead to a delay in diagnosis of a bleeding disorder related to factor ix deficiency. this may lead to additional diagnostic testing and delayed prophylactic therapy.
  • 조치
    Siemens has provided a workaround for users along with a technical support contact number, whilst a permanent fix is being developed. Customers are advised to follow their laboratory protocol regarding the need for review of previously reported patient results. This action has been closed-out on 01/03/2018.