FilmArray Blood Culture Identification (BCID) Panel when used with BD BACTEC Blood Culture Bottles. An in vitro diagnostic medical device. 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Biomerieux Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00271-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-04-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Biomérieux australia has been advised that the manufacturer, biofire diagnostics, has identified an increased risk of false positive proteus results when the filmarray blood culture identification (bcid) panel is used with bd bactec blood culture bottles with expiration dates of september 30, 2018 and october 31, 2018. it is unknown if subsequent lots of media will also be subject to this risk; however, biofire and bd are continuing to investigate the issue.
  • 조치
    bioMérieux is advising customers to verify positive results for Proteus by confirming with another method where BD BACTEC Blood Culture Bottles with expirations on 30/09/2018 and 31/10/2018 have been used due to the BCID Panel detecting nucleic acid from non-viable Proteus.

Device

  • 모델명 / 제조번호(시리얼번호)
    FilmArray Blood Culture Identification (BCID) Panel when used with BD BACTEC Blood Culture Bottles. An in vitro diagnostic medical device. All lots of FilmArray Blood Culture Identification (BCID) Panel are impacted when used with specific BD media typesMultiple BD BACTEC Blood Culture Bottles with expiry dates of September 30 2018 and October 31 2018 may impact resultsARTG Number: 223979(Biomerieux Australia - Bacterial infectious disease IVDs)
  • Manufacturer

Manufacturer