Flexi-Seal CONTROL Faecal Management System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Convatec Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00507-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-05-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An internal assessment of product performance, including a review of customer feedback, has confirmed that flexi-seal control faecal management system is not meeting performance requirements for this product. specifically, the auto-valve has not been consistently performing relative to the inflation and deflation of the device’s retention balloon. in the event that the auto-valve does not perform properly related to inflation and deflation of the device’s retention balloon, there is the potential that medical intervention may be necessary. in accordance with the instructions for use supplied with the flexi-sealtm controltm faecal management system, the following events could occur:- leakage of stool around the device, and peri-anal skin breakdown;- rectal/anal bleeding due to pressure necrosis or ulceration of rectal or anal mucosa;- perforation of the bowel.
  • 조치
    End users are advised to stop the use of all affected devices and return all affected products to their distributor for credit or replacement with unaffected product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Flexi-Seal CONTROL Faecal Management SystemProduct code: 411107All lot numbersARTG number: 163198
  • Manufacturer

Manufacturer