Flexi-Slip endotracheal tube stylets(Intended for use in facilitating intubation of tracheal tubes.) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Teleflex Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00874-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-09-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There have been reports of the plastic coating of the stylet splitting and/or breaking off of the stylet. this may result in a piece of plastic totally or partially occluding the patient’s airway and impairing ventilation, or necessitating invasive removal procedures in order to prevent complications such as atelectasis or pneumonia. over 200,000 units have been distributed and there have been 4 complaints to august 2015. this gives rise to a complaint rate of 0.00205%.No patient injuries have been reported related to this issue.
  • 조치
    Customers are asked to inspect their stock, quarantine any affected units from use and to follow the instructions on the acknowledgement form in order to have those units collected. An appropriate account credit will then be issued. This action has been closed-out on 22/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Flexi-Slip endotracheal tube stylets(Intended for use in facilitating intubation of tracheal tubes.)SAP/Catalogue Number: 502501-000060Batch Numbers: 14BG27, 14FG03ARTG Number: 214058
  • Manufacturer

Manufacturer