Flexible Uretero-Reno Videoscope & Flexible Uretero-Reno Fiberscope 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Olympus Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01519-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-12-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Omsc has received a small number of adverse event reports globally for the urf-v2 associated with damage to the bending section. to date, no adverse events have been reported to olympus in australia for the urf-v2 and no adverse events reported globally for the urf-p6. further to olympus’ safety alert issued in december 2016 (rc-2016-rn-01653-1), the manufacturer has modified the design of the bending section of the urf-v2 and urf-p6 to address any deficiencies that may have attributed to the adverse events reported globally.
  • 조치
    Olympus is advising hospitals to identify affected devices from their inventory. When replacement URF-V2 and URV-P6 devices are available, an Olympus Customer Operations Specialist will contact users to exchange devices at affected facilities. In the interim, users may refer to the IFU and Safety Alert issued by Olympus for the correct inspection and use of the device.

Device

  • 모델명 / 제조번호(시리얼번호)
    Flexible Uretero-Reno Videoscope & Flexible Uretero-Reno Fiberscope Model Numbers: URF-V2 and URF-P6ARTG Numbers: 211118 and 131068
  • Manufacturer

Manufacturer