Flocare Gastrotomy Tubes (G-tube) ENFit connector 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Nutricia Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00725-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-06-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Nutricia has received reports that the connector from the flocare g-tube enfit will, in in some occasions, detach. this situation generally occurs while disconnecting the enfit g-tube from the enfit feeding set or enfit syringe. if the connector is separated from the g-tube enfit, leakage of nutrition or spillage may occur resulting in:- an interruption or delay in therapy. when this situation remains unnoticed for a prolonged period of time, the health of a volume dependent patient can be affected due to under feeding.- a risk of abdominal burns caused by spillage of gastric contents.
  • 조치
    Nutrica is advising users that in order to minimise the risk of separation, they are to 1) avoid the application of excessive force or overtighten the ENfit connectors, 2) keep the ENFit connectors as clean as possible from residual nutrition and 3) flush/clean the threads of the ENFit connectors routinely to remove residual nutrition. In the event of separation, users are advised to close the quick release clamp immediately and request a suitable adaptor from Nutricia to use in place of the ENFit end. The Medicina Trainsition Adaptor will be provided and users are advised to refer to their recall letter for its proper use.

Device

  • 모델명 / 제조번호(시리얼번호)
    Flocare Gastrotomy Tubes (G-tube) ENFit connectorFlocare Gastrostomy tube (G-tube) Ch10 - ENFitSKU: 594814Flocare Gastrostomy tube (G-tube) Ch14 - ENFitSKU: 594815Flocare Gastrostomy tube (G-tube) Ch16 - ENFitSKU: 594816Flocare Gastrostomy tube (G-tube) Ch18 - ENFitSKU: 594817Flocare Gastrostomy tube (G-tube) Ch20 - ENFitSKU: 594818All BatchesARTG Number: 120000