Flocare Infinity Pump and Infinity Plus Pump (enteral feeding pumps) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Nutricia Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01050-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Nutricia australia has received reports from overseas where the air in line (ail) alarm functionality of the flocare infinity enteral feeding pump did not work as expected when used with a specialised modular feeding regimen.The air in line (ail) alarm functionality may not always work with mixed or reconstituted powdered nutrition. un-dissolved particles may accumulate in the enteral feeding line or form a film on the inner side of the tube, around the pump sensors, which may interfere with pump air detection algorithms. this may result in the delivery of (excess) air into the patient. if the ail alarm is used as an indicator that the reservoir is empty (therapy completed), follow up therapy may be delayed and may place critical care and volume sensitive patients at risk.To date, no ail alarm issues have been identified with the use of ready-made liquid feed tube nutrition.Nutricia is liaising with the tga to ascertain if further action needs to be undertaken.
  • 조치
    Nutricia is advising that for patients where interruptions or delay in therapy could affect their health status, the AIL alarm should not be used to indicate end of dose and the DOSE setting of the Infinity pump must be used. It should be ensured that the volume of nutrition in the container is greater than the volume that is actually needed, i.e. more than the set DOSE to be administered, so as to avoid excessive air bubbles getting in the feeding line. Nutricia is also advising users to ensure mixed or reconstituted powdered nutrition is prepared and dissolved properly to avoid particles or accumulation in the feeding set potentially hampering pump performance and a normal feeding regimen. Any nutrition added into the enteral feeding line must be of a homogeneous nature and should remain homogeneous. Users must ensure that any substrate used is indicated as suitable for enteral tube feeding delivery (refer to instructions for use on nutrition / substrate added). Nutricia recommends that, where possible, patients use ready-made liquid tube nutrition, as this is developed for the delivery specifications of the pump. For more information, please see https://www.tga.gov.au/alert/flocare-infinity-enteral-feeding-pumps . This action has been closed-out on 18/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Flocare Infinity Pump and Infinity Plus Pump (enteral feeding pumps)Product Codes: 35679, 35680All Batch Numbers affectedARTG Number: 119667