Flocare Transition Giving Sets (Enteral feeding tube) Flocare Infinity Pack Giving Set MLL W/O DC – TransitionFlocare Infinity Bottle Giving Set – TransitionFlocare Infinity Pack Giving Set W/O DC & W/O MP - TransitionFlocare Infinity Bottle Giving Set W/O MP – TransitionFlocare Infinity Pack Mobile Giving Set – TransitionFlocare Infinity Pack Mobile Giving Set W/O MP - Transition 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Nutricia Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00063-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-01-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Some users overseas have experienced issues with leakage and/or breakage of the enfit (white) transition adaptor included in the flocare transition giving sets (mll and mobile giving sets only). leakage could occur immediately after connection to a luer feeding tube and initiation of the feeding regime, as well as over time.The white transition adaptor may only be required by a small number of patients in australia to connect to a luer feeding tube.Higher volumes of leakage (spoilage of critical nutrition) might lead to an under infusion situation which, especially for critical and volume sensitive patients, can have a negative impact on patient health if it remains unnoticed for a longer period of time.
  • 조치
    Users who are using the white transition adaptor are advised that they can continue to use the affected products. However, when using, they are advised to check giving set adaptors for any minor defects, ensuring that connectors are not over-tightened which may cause cracking, and monitoring for leakage. Furthermore, as a precautionary measure, patients/parents/carers are advised to check for any leakages after initial connection and again after approximately 2 hours. If there are signs of leakage please change the set for a new one. It is anticipated that improved adaptors will be available in the market as of April 2016. For further information, please see http://www.tga.gov.au/alert/flocare-transition-giving-sets-enfit-transition-adaptors .

Device

  • 모델명 / 제조번호(시리얼번호)
    Flocare Transition Giving Sets (Enteral feeding tube)Flocare Infinity Pack Giving Set MLL W/O DC – TransitionFlocare Infinity Bottle Giving Set – TransitionFlocare Infinity Pack Giving Set W/O DC & W/O MP - TransitionFlocare Infinity Bottle Giving Set W/O MP – TransitionFlocare Infinity Pack Mobile Giving Set – TransitionFlocare Infinity Pack Mobile Giving Set W/O MP - TransitionCodes: 595168, 595172, 595174, 595179, 595181, 595182All batch numbers affectedARTG #: 119666
  • Manufacturer

Manufacturer