FLOSEAL Special Applicator Tips 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Baxter Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00670-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Baxter healthcare has identified an issue with the use of an incorrect plastic formulation in the manufacture of the blue luer component of the floseal malleable tips. the plastic formulation does not present any toxicity or biocompatibility concerns, but does impact the physical characteristics of the blue luer component. the affected lots of the floseal malleable applicator tip have been associated with the cracking of the luer upon connection with the syringe. since the tip is attached to the syringe outside of the surgical field, it is unlikely that plastic fragments would reach the surgical field. nevertheless, in exceptional cases, detached subvisible particles may be retained in the body and may cause inflammation, or serve as a nidus for bacteria in already contaminated wounds and contribute to allergic reactions. to date, there have been no reports of serious injury associated with this issue.
  • 조치
    Baxter is advising customers to locate and return all units of the affected product. Instructions for return can be found on the Customer Letter.

Device

  • 모델명 / 제조번호(시리얼번호)
    FLOSEAL Special Applicator TipsProduct Code: 1502186Box Lot Number: HA170818 Pouch Lot Numbers: GR338160 (expiry 31 May 2019) and GR338384 (expiry 30 June 2019)ARTG Number: 167776(Baxter Healthcare - Haematological concentrate system applicator, endoscopic)