Fluorocell PLT reagent. An in vitro diagnostic medical device (IVD)(used on Sysmex XN-10 and XN-20 Automated Haematology Analysers) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01275-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-10-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Sysmex has advised that specific lots of the fluorocell plt reagent may be susceptible to environmental factors causing a decrease in their fluorescence over time. in such instances, the affected lots could lead to a false low plt-f value, which will be inconsistent with the plt-i value.
  • 조치
    Roche is advising customers to discard any affected lots of Fluorocell PLT in use or stock. If available, users should use cartridges of an unaffected lot of Fluorocell PLT. Users should ensure that they perform reagent replenishment for Fluorocell PLT to flush out the old lot and completely replenish with unaffected reagent. If no other lot of Fluorocell PLT is available, please check PLT-F results for plausibility. If the PLT-F results are not plausible, please refer to the PLT-I results from the impedance channel. Users should follow internal SOP's regarding retrospective judgement of affected samples. Contact Roche Diagnostics Australia to discuss the need for replacement product.