Fujifilm Synapse PACS software version 4.0, 4.1, 4.2, 4.3 (a picture archiving and communication system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Fujifilm Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01329-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-11-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An issue has been identified with the image processing software module when two or more patient fcr images (fuji computed radiology images stored in a proprietry format) from synapse pacs are sent simultaneously to another manufacturer's pacs . when fcr images (comprised of pixel data and the dicom header) are sent via cmove or forwarded by ebf (event based forwarding) on multiple associations simultaneously, in rare situations, one or more of the following issues can happen: a) patient a’s pixel data is changed to patient b’s pixel data and forwarded to the destination. b) the output image is corrupted and may appear as random noise. c) dicomserver crashes. if issue a) or b) happens, only the image in the destination archive is affected. images in the source synapse system are not affected. the potential risk is that pixel data for a patient’s image may not be correct in the destination (non-fujifilm) pacs. however, the radiology reports will be correct.
  • 조치
    Fujifilm has applied a corrective software patch to affected PACS as a permanent fix. This action has been closed-out on 20/11/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Fujifilm Synapse PACS software version 4.0, 4.1, 4.2, 4.3 (a picture archiving and communication system)ARTG Number: 203605
  • Manufacturer

Manufacturer