GE Healthcare Revolution EVO & Optima CT Systems CT660, CT670, CT680, CT540 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00499-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-05-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has identified that some ct systems may have a damaged cable that can expose 120vac to service engineers working inside the gantry. this issue does not affect patients or operators external to the equipment.When the gantry front cover is removed to allow servicing of components inside the gantry, cover dollies are used to manipulate the cover. during the removal or reattachment of the front cover, a bracket can contact an internal power supply cable and damage that cable’s insulation. this can expose the conductor of a 120vac power supply cable and electrify the cover dolly, thereby introducing the risk of an electrical shock hazard to the service personnel.
  • 조치
    GE Healthcare is advising they will be in contact with affected users to arrange for the correction to be implemented. In the interim, the system may continue to be used as this issue does not impact patients or CT operators. Users should ensure service technicians are aware of the issue and to wear protective gloves when using the cover dollies. GE Healthcare have updated the service manual to prevent recurrence of this issue in the event of a replacement of the cable in the future. Users should follow the instructions outlined on the Customer Letter to download the latest Service Manual.

Device

  • 모델명 / 제조번호(시리얼번호)
    GE Healthcare Revolution EVO & Optima CT SystemsCT660, CT670, CT680, CT540Revolution EVO Model Numbers: 5454001-61, 5454001-160, 5454001-60 and 5454001-260 Optima CT660Model Numbers: 5454001, 5454001-41, -100, -220, -200, -40, -22, -140, -240, -122, -222Optima CT670/680 Model Numbers: 5454001-230, 5454001-330 Optima CT540Model Numbers: 5432539, 5432539-2, 5432539-3, 5432539-5 and 6447929ARTG Number: 116979