Getinge 46-Series Washer Disinfector with PACS 300/350 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getinge Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00174-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-03-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has received complaints in germany regarding sediment residuals in the manifold of the washer and disinfector of getinge 46-series (cm302) with pacs 300/350 control.The potential risk is that during the drying phase the air circulating may transport particles from the sedimentation and deposit it on goods. if the particle contains bacterial flora there is the risk of cross contamination.The problem has been identified as insufficient water flow throughout the manifold due to formation of air-pockets where sedimentation may occur. the use of detergents that are not recommended might also have negative impact.No serious injury has been reported in any of the known cases.
  • 조치
    Getinge will contact customers to carry out a modification that will permanently mitigate the risk of this issue occurring in the future. In the interim customers are advised to perform a visual check of all washed items after removal from the washer. Items should be re-washed if any particles are observed. This action has been closed-out on 1/06/2016.