GF-UCT180 Ultrasonic Gastrovideoscope 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Olympus Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00909-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-09-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The nature of this product correction is to issue a revised instruction for use (ifu) which includes instructions on how to correctly clean the device. there is a risk of patient-to-patient infection if the device is not cleaned appropriately. key changes in the new instruction for use are highlighted below:· inclusion of instruction on how to attach the cleaning adapter (maj-350) to the instrument channel port of the gf-uct180· the balloon applicator (maj-675) and the cleaning brush (maj-1534) are to be reprocessed for reuse. the maj-675 and maj-1534 are to be reprocessed in accordance with step “cleaning, disinfection, and sterilisation procedures for reusable parts and reprocessing equipment”.
  • 조치
    Olympus is advising their customers of the amendments to the IFU. Customers can request a new IFU by contacting Olympus. Hospitals are advised that a review of internal hospital protocols should be conducted to ensure compliance with the updated cleaning instructions. This action has been closed-out on 29/08/2016.