Glucose-6-Phosphate Dehydrogenase. An in vitro diagnostic medical device (IVD).(quantitative kit used for determination of G-6-PDH in blood) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Immuno Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00081-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-01-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, trinity biotech, has identified a potential for misdiagnosis with lot f117009. an investigation identified that this lot displays a level of accuracy variation outside the historical qc product release process control charts for normal range controls, despite falling within the qc acceptance specification. investigation testing has shown that there is objective evidence that the accuracy issue affects the normal patient range and not the deficient range with a potential health hazard of a false positive result (normal patient diagnosed as deficient) or an invalid test result occurring.
  • 조치
    Immuno is advising customers to immediately discontinue use of the affected lot and destroy any remaining product. Laboratories are to consider the need to review results previously generated with the affected lot. Immuno will organise replacement kits.