Greatbatch Medical Offset Reamer Handle (used in total hip arthroplasty as interface between surgical drill and acetabular reamer) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Biomet Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01593-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Greatbatch medical has identified an issue with the 00-7804-080-00 offset reamer handle, as field complaints were reported that the drive chain mechanism seizes during use. the function of the drive chain mechanism is to transmit rotational power from a mechanical drill to an acetabular reamer during a total hip arthroplasty. during seizing, the device is rendered non-functional in the operating room. the risks associated with this failure are: 1. the offset reamer handle seizes and the surgery must be completed with an alternate device, leading to a delay of 30 minutes or greater from the scheduled surgery time. 2. the offset reamer handle seizes and no replacement device is available in the hospital, leading to an interruption and rescheduling of surgery.
  • 조치
    Zimmer is advising that a re-designed Offset Reamer Handle will be available in early 2017 and that customers will be notified once the replacements are available for delivery by Zimmer. In the interim, Zimmer is advising that a Straight Reamer Handle should be always be available and ready for surgical use and/or that a replacement or alternate Offset Reamer Handle, if desired, also be ready for use in a surgical procedure in or near the operating room. A Straight Reamer Handle device is commonly provided as part of the Reamer surgical set. This action has been closed-out on 28/02/2018.

Device

  • 모델명 / 제조번호(시리얼번호)
    Greatbatch Medical Offset Reamer Handle (used in total hip arthroplasty as interface between surgical drill and acetabular reamer)Part Number: 00-7804-080-00Lot Numbers: 56474941, 56474943, 56474944, 56474947, 56474982, 56474983, 56474990, 56596272, 56596277, 56596292, 56596366, 56596371, 56596428ARTG Number: 266809
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA