Guardian Connect Mobile App 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00882-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-07-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Medtronic has identified an issue in apple ios 11.3 & 11.4 that can sometimes prevent patients from receiving a 'lost communication' alert after their guardian connect app is closed by the iphone's operating system (ios) in certain situations. the operating system may close the guardian connect app without alerting patients that the app is no longer running or communicating with the guardian connect transmitter. if the app is closed, patients will not receive alerts that could be associated with hypoglycaemic or hyperglycaemic events. medtronic have received reports from patients regarding the app closing without warning, however, to date, no reports of injury as a result of this issue have been received.
  • 조치
    Medtronic is advising users to immediately update their Guardian Connect App to the latest release (v3.2.2) located on the App Store. Medtronic also reminds users that as described in the App User Guide, Medtronic cannot guarantee that iPhones will always allow for the Guardian Connect App to run in the background. If the iPhone automatically closes your Guardian Connect App, users may receive a ‘Lost Communication’ alert. To ensure users continue receiving sensor glucose data and alerts, Medtronic recommends periodically checking that the Guardian Connect App is still running, being especially mindful of this anomaly when running several applications at once or after using gaming and video Apps which use a lot of the iPhone’s power or memory resources.

Device

  • 모델명 / 제조번호(시리얼번호)
    Guardian Connect Mobile AppPart number: CSS7200 ARTG Number : 276924(Medtronic Australasia Pty Ltd - Information system software, application program, patient record)
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA