Guide Sleeve, yellow for Trochanteric Fixation Nail Advanced (TFNA) Proximal Femoral Nailing System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00877-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The affected lots of the guide sleeves (part # 03.037.017) may have been manufactured with a slightly larger than specified outer diameter which could prevent the sleeve from going through the trochanteric fixation nail advanced (tfna) aiming arm properly.In the event that guide sleeve does not fit through the tfna aiming arm properly, a surgical delay may occur.
  • 조치
    JJM DePuy Synthes is advising users to locate any potentially affected devices in their facility, and contact their JJM DePuy Synthes representative. The representative will conduct functional testing on all products subject to this notification to determine whether the device is conforming or not. Users should ensure every aiming arm is tested with every guide sleeve within their control and, upon completion, only return the device if the product does not pass the functional test.

Device

  • 모델명 / 제조번호(시리얼번호)
    Guide Sleeve, yellow for Trochanteric Fixation Nail Advanced (TFNA) Proximal Femoral Nailing SystemPart Number: 03.037.017Lot Numbers: 9066435, 9469298, 9542486, 9734863, 9763550, 9768881, 9768882, 9768884, 9768886ARTG Number: 153950