Hallu-Fix Drill, Diameter 1.9mm, supplied in Hallu-Fix system instrumentation set (used for orthopaedic foot surgery) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Integra Neurosciences Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01387-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-10-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has identified a defect in lot fe58 of hallu-fix drill diameter 1.9mm, where the drill may not be sharp enough to prepare screw holes. the defect can be easily recognised during use and an increase of surgery time may be observed while another similar drill is found to prepare the screw holes. if an affected drill is used, the defect can contribute to an increase of surgery time of approximately 10 minutes and require a higher pressure on the drill. this pressure may cause minimal transient tissue damage and difficulties to achieve the surgery. to date no injury or adverse event has been reported.
  • 조치
    Integra Neurosciences is advising customers to review their inventory for the affected drill and stop using the affected drill immediately. Integra will provide directions for return of the product and will replace affected stock. This action has been closed-out on 18/05/2017.