Hamilton-G5 and Hamilton -S1 Ventilators with software versions v2.41, v2.42 and v2.50 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 AHC Solutions Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00450-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The analysis of a customer complaint has identified an issue related to suctioning with the use of the hamilton-g5 and hamil ton-s1 ventilators. after performing the suctioning manoeuvre, including disconnecting the patient, suctioning, and then reconnecting the patient, the preset pattern of ventilation may not continue as expected. under certain conditions, a different ventilation mode than the operator selected mode maybe applied; this situation can occur independent from the selected patient group (neonate, paediatric, or adult). this situation may occur if the ventilation mode changes during the pre-oxygenation phase. after the suctioning manoeuvre is finished and the patient is reconnected, the ventilation mode active before the pre-oxygenation phase is used. under these conditions, the ventilation mode displayed by the ventilator is not the same as the applied ventilation mode. in the worst case, the patient may be either hypo ventilated or hyperventilated.
  • 조치
    AHC is providing instructions to deactivate the suctioning option on all affected ventilators until a software upgrade to version 2.60 has been installed. Users are advised to ensure there are alternative suction devices available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Hamilton-G5 and Hamilton -S1 Ventilators with software versions v2.41, v2.42 and v2.50Product codes: 159001 & 159005Multiple Serial numbers affected ARTG number: 113999
  • Manufacturer

Manufacturer