Hamilton Medical MR1 Ventilator 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01133-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An issue has been identified where the oxygen tubing and the oxygen connector of the hamilton-mr1 became loose during the preparation for ventilation. in the case of loose oxygen tubing during operation, the ventilation is continued and the “oxygen supply failed” alarm is triggered if the oxygen setting is above 21%. the oxygen connector may hurt the patient and the operator due to the high pressure of the oxygen and uncontrolled movement of the tubing. there have been no injuries as a result of this issue.
  • 조치
    Taleb Medical is advising users to check the oxygen connector screws for tightness as an immediate measure. Tighten with a screwdriver if necessary. Do not over tighten the screws as this can damage the thread. If the oxygen connector cannot be tightened or if the screw-hole threads are damaged, do not use the ventilator. A new O2 mixer mounting plate will be installed as a permanent correction.