HARMONIC ACE + Shears with Adaptive Tissue Technology 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00446-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-05-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Through post-market surveillance efforts and through an internal investigation, the manufacturer ethicon, have confirmed that some devices contained in these lots may have been assembled with an internal component that may cause continuous or inadvertent activation of the device.To date, ethicon has not received any reports of adverse events associated with this issue.No other lots of the harmonic ace+ devices (other than those listed) and no other harmonic products are affected by this action.
  • 조치
    Johnson & Johnson are advising customers to inspect devices in their possession to determine if their facility has affected product. Customers should then follow the instructions provided in the Acknowledgement Form provided with the Customer Letter to arrange return of affected product.

Device

  • 모델명 / 제조번호(시리얼번호)
    HARMONIC ACE + Shears with Adaptive Tissue TechnologyProduct Code: HAR23 and HAR231Lot Numbers: P9125C, P94C8R, P94G1W, P9125C, P9125C, P94C8R, P94G1W, P9125CARTG Number: 118011(Johnson & Johnson Medical - Electrode, electrosurgical, active, ultrasonic/mechanical vibration, single-use)
  • 의료기기 분류등급
  • Manufacturer

Manufacturer