HeartWare Ventricular Assist System (intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end stage heart failure) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Heartware Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01323-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-12-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Subsequent to a urgent recall for product correction (tga ref: rc-2013-rn-00126-1) of hvad pump's driveline connector housing become partially or fully separated from the front portion of the driveline connector, heartware have identified an alternative failure mechanism. in 8 cases, customers have reported that the locking mechanism of the driveline connector has failed to engage. there have been no reports of injury. heartware has made changes to the manufacturing procedures to address this issue.
  • 조치
    Heartware is advising users that should the driveline disconnect from the controller, push the driveline connector back into the controller immediately and contact Heartware for repair.. Heartware is advising clinicians to inspect patient's driveline connector as described in the provided technical bulletin for proper locking. For already implanted patients, please arrange a follow up visit at the earliest convenience to check the patient's driveline connector. If the locking mechanism of the driveline connector is found to have failed to engage, please push the connector back into the controller and contact Heartware for repair.

Device

  • 모델명 / 제조번호(시리얼번호)
    HeartWare Ventricular Assist System (intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end stage heart failure)Catalogue Numbers: 1100, 1101, 1102, 1103, 1104 & 1205Serial Numbers: HW001 to HW11215 and HW20001 to HW20296ARTG Number: 181875
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA