HexaPOD evo RT Systems running iGUIDE 2.0.0 - 2.0.2 (used for accurate patient positioning within a radiation therapy treatment environment) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00612-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-06-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The iguide key allows to the treatment of patients without using the hexapod evo rt system. when users turn the iguide key to the off position; the external inhibit signal is overridden and the iguide software closes any open fractions and logs off (login appears). in some circumstances, a bug in the iguide software prevents the iguide software from logging off. the iguide software looks like it can be used for positioning, but it is no longer running correctly. even when users turn the iguide key back to the on position, it will not set the external inhibit signal or alert them if the position has changed (ie, position guard is inactive). this problem can lead to clinical mistreatment if the user assumes that the iguide software is running correctly. the iguide software will not set the external inhibit signal or notify the user in case of incorrect positioning or movements.
  • 조치
    Users should only use the iGUIDE software when the key is in the on position and the Position Guard is active. Users are advised to restart the workstation if any error occurs to ensure the software is running properly. A software update will be implemented on all affected systems that corrects the behaviour of the iGUIDE software. This action has been closed-out on 11/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    HexaPOD evo RT Systems running iGUIDE 2.0.0 - 2.0.2 (used for accurate patient positioning within a radiation therapy treatment environment)ARTG Number: 187340