HomeChoice PRO Automated Peritoneal Dialysis (PD) System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Baxter Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00304-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-03-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Baxter healthcare is implementing additional warning in homechoice/homechoice pro device product labelling related to patients more vulnerable to unintended increased intraperitoneal volume (iipv), also known as overfill. specific populations using peritoneal dialysis (pd) therapy including neonates, infants, small children and any patient with significant heart or lung disease are more vulnerable to the negative effects of excessive retention of fluid in the peritoneal cavity that may be related to their dialysis therapy. overfilling with dialysate (iipv) or not draining enough fluid during pd therapy can result in excess fluid in the abdomen accompanied by patient symptoms.
  • 조치
    The additional warning identifies symptoms related to unintended IIPV that may occur in specific patient populations and provides specific instructions to the patient and/ or caregiver to follow if IIPV is suspected. For more details, please see http://www.tga.gov.au/safety/alerts-device-automated-peritoneal-dialysis-system-140318.htm .