Horizon Cardiology Hemo1 – versions 12.2 HF4 and earlier(used for acquiring, displaying and reviewing haemodynamic parameters (vital signs) in cardiac catheterisation laboratories) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00735-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-07-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Mckesson has identified that its communication to customers regarding the required maintenance tests was insufficient. while each unit has an attached sticker indicating the date of the next maintenance test the need for performing annual maintenance testing is not reiterated in 12.2 hf4 and earlier versions of the horizon cardiology hemo user guides. if the parameter boxes are not calibrated, the horizon cardiology hemo system may not function properly or potentially affect the accuracy of clinical parameters.
  • 조치
    Users are advised that the new version of the User Guide (12.2 version 5.0) is now available to address this issue and will be provided electronically upon request. The attachment provided with the customer letter identifies the updated sections of the new User Guide related to annual maintenance testing. This action has been closed-out on 02/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Horizon Cardiology Hemo1 – versions 12.2 HF4 and earlier(used for acquiring, displaying and reviewing haemodynamic parameters (vital signs) in cardiac catheterisation laboratories)ARTG number: 146573
  • Manufacturer

Manufacturer