Horizon Medical Imaging (HMI) products software versions 4.5 to and including 11.9 (Radiology picture archiving and communication system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00101-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-01-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During the archive procedure, if the system begins to archive images for multiple studies within fractions of a second and an unexpected condition causes the software process restart the following issues may occur:1. incorrect images being displayed for a study retrieved from archive (for example, a study for a different patient being displayed),2. failure to retrieve images for a study from archive.If this issue occurs, two or more studies may be affected. this issue is not immediately evident to the user at the time of archival. it is possible that clinical decisions will be made using an incorrect set of images or an incomplete set of images. this may result in inappropriate care leading to patient harm if unnoticed or a need to retake images.
  • 조치
    Device Technologies is providing users with work around instructions to follow as an interim measure. A software upgrade will be implemented as a permanent correction.

Device

  • 모델명 / 제조번호(시리얼번호)
    Horizon Medical Imaging (HMI) products software versions 4.5 to and including 11.9 (Radiology picture archiving and communication system)ARTG Number: 158895