HOYA Corporation (PENTAX Medical) – Gas/Water Feeding Valve 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00711-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-07-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This action is taken to inform users of a change in the instructions for use related to the proper use of the lubricant oil (ptx-of-z11) applied on the ptx-of-b130 and ptx-of-b194 gas/water feeding valves, as supplied by pentax medical.The pentax brand silicon oil ptx-of-z11 must always be used for the o-rings of pentax gas/water feeding valves. at this time, pentax medical can only recommend ptx-of-z11 as it is the only lubricant validated by pentax medical for use with these valves. if other oil were applied to the o-ring of ptx-of-b194 this may impact the functionality, and it may cause uncontrolled, continuous gas flow or excessive gas insufflations, resulting in a patient injury such as pneumatic perforation or gas (air) embolism.
  • 조치
    Device Technologies is providing updated Instructions for Use with the customer letter. Users are also advised that If the valve does not function properly (does not move smoothly, feels sticky, leaks air or water) remove it and inspect for worn or damaged O-Rings. Replace O-rings as necessary. In case the valve does not function properly during an examination, stop CO2 or water supply at the primary valve (i.e CO2 insufflator or CO2 bombe) immediately.

Device

  • 모델명 / 제조번호(시리얼번호)
    HOYA Corporation (PENTAX Medical) – Gas/Water Feeding ValveProduct Codes: PTX-OF-B130 & PTX-OF-B194ARTG Number: 204881