HS66 Turbo LTSF Steriliser 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getinge Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00908-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-08-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The problem identified indicates a leakage at a critical valve due to clogging of solidified formalin solution, also known as paraformaldehyde. this leakage might affect the efficacy of the ltsf sterilisation cycle and cause potential non-sterility. the problem appears stealthy until clogging has reached the level where the built-in leakage monitoring system of the steriliser is able to indicate leakage and cause an alarm. the potential risk caused by clogging is that sterile filtered air can be introduced into the sterilizer chamber during the sterilization cycle and may partially block penetration of formaldehyde at items subjected to sterilization. the risk for non-sterility is assessed to increase at hollow instruments compared to solid instruments.
  • 조치
    A Getinge representative will schedule a service technician visit to inspect, clean and take preventive measures at your sterilizer for the problem not to reappear. In the meantime all users are advised to take precautions as follows: - Depending of the occurrence and frequency of LTSF sterilization at your sterilization department use biological indicators for monitoring the sterilization efficacy at a daily or weekly basis; - When growth is indicated by biological indicators - stop using the sterilizer and contact your Getinge representative for prompt assistance. This action has been closed-out on 03/06/2016.