iChem Velocity Urine Chemistry Strips. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00806-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-09-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, iris international has determined that sample probe misalignment or bending on the ichemvelocity may lead to a remote possibility of false negative results due to inadequate strip dosing. the manufacturer'sinvestigation was initiated based on customer complaints for control failures. iris international has become aware of a manufacturing defect where a glucose pad was placed in the location intended for the protein pad on a subset of vials numbered 14996 through 15859 (with the exception of vials 15315 and 15316 within that series) within lot number 7204081a.The manufacturing defect will cause:· consistent false negative results in the presence of proteinuria.· false positive urine protein results may occur when patient specimen contains glucose.· quality control failures for protein results.
  • 조치
    Beckman Coulter is advising users to inspect and discard any remaining inventory of affected vials. Beckman Coulter recommends consulting with the Laboratory Director to determine if a review of previously reported results is required. This action has been closed-out on 18/08/2016.