iChemVelocity Automated Urine Chemistry System using iChemVelocity Urine Chemistry Strips. An in vitro diagnostic device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00065-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-02-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There have been complaints indicating that loose or missing analyte pads being found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the ichemvelocity system.If affected strips are used to assay patient samples, the issue has the potential to cause erroneous ph results and false negative results for the following tests:nitrite, glucose, blood, bilirubin, urobilinogen, protein, leukocyte esterase, ketones, and ascorbic acid.There is no impact to the measurement of colour, clarity and specific gravity.
  • 조치
    The sponsor is advising users to discard the vial and not use the strips from that vial when loading strips if loose or missing analyte pads are observed. Users should contact the sponsor for product replacement. A review of patients' results is at the discretion of the laboratory director. This action has been closed-out on 16/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    iChemVelocity Automated Urine Chemistry System using iChemVelocity Urine Chemistry Strips. An in vitro diagnostic device (IVD)Reference Numbers: 800-7212; 800-7212-001; 800-7204; 800-7204-FJRAll Lot and Serial Numbers are affectedARTG Number: 177551