iChemVELOCITY Automated Urine Chemistry Systems equipped with Colour/Clarity/Specific Gravity Module. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00477-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-04-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The ichemvelocity colour/clarity/specific gravity module (cgm) detects and reports 11 colours as opposed to the 18 listed in the ichemvelocity operator’s manual. if the ichemvelocity automated urine chemistry system is equipped with any of the affected cgm part numbers, the following urine colours will be detected and correctly reported: colourless, straw, light yellow, yellow, dark yellow, light amber, amber, dark amber, red, blue, and green. the following urine colours will not be detected or reported: orange, light brown, brown, dark brown, light red, dark red, other. there may be a discordance between the colour of the urine that is visually apparent and the colour that is reported, for the affected colours. in a worst case scenario, a discordant urine colour might prompt an additional non-invasive workup such as a repeat urinalysis or further testing. the results from the urine chemistry and microscopy will still be correctly reported.
  • 조치
    Beckman Coulter is advising users that the iChemVELOCITY Operator’s Manual will be updated to reflect colours that can be detected and reported by the affected CGMs. Furthermore, if the urine colour orange, light brown, brown, dark brown, light red, dark red or other colour than those listed as ‘can be detected and reported’ is visually observed during processing, it is recommended that users: - Consider reporting the visually observed colour reading using the laboratory’s reporting system (e.g. hardcopy, LIS). - Consider using an approved back-up analyser to process and report these samples if the colours above are needed within the laboratory.

Device

  • 모델명 / 제조번호(시리얼번호)
    iChemVELOCITY Automated Urine Chemistry Systems equipped with Colour/Clarity/Specific Gravity Module. An in vitro diagnostic medical device (IVD)System Reference Numbers: 800-3564, 800-7100, 800-7103, 800-7162, 800-7163, 800-7166, 800-3061Colour/Clarity/Specific Gravity Module Reference Numbers; 700-7602, 700-7611, B93997ARTG Number: 181116