iChemVELOCITY Urine Chemistry Strips 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01422-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-11-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Iris international has become aware of a strip manufacturing error where an incorrect pad was inadvertently placed in the location of the compensation pad. all the other analytes are correct. this defect exists on a subset of vials belonging to lot 7204170 a with the serial numbers of the impacted vials ranging from 10500 through 12818. beckman coulter is working on a resolution to prevent recurrence of this issue. this action is being taken following consultation with the therapeutic goods administration.
  • 조치
    1. Inspect any remaining inventory to determine if you have the impacted vials The vial serial number is located on the bottom of the vial beneath the lot number. 2. Discard any remaining impacted vials. 3. Contact your local support representative for product replacement. 4. Consult with your Laboratory Director to determine if a retrospective review of results is clinically warranted. 5. Share this information with your laboratory staff and retain this notification. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 6. Complete and return the supplied Response Form within 10 days so Beckman Coutler can be assured you have received this communication.

Device

  • 모델명 / 제조번호(시리얼번호)
    iChemVELOCITY Urine Chemistry StripsLot number: 7204170 ASerial Nunber: 10500 through 12818ARTG Number: 177551
  • Manufacturer

Manufacturer