Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos RHD, Luminos RF Classic, Luminos Fusion, Luminos Select (Fluoroscopic diagnostic x-ray systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00111-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-02-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is potential for malfunction in cases where the table vertical tilt movement is obstructed by a solid object resulting in a collision. this can result in damage to the system mechanics that may not necessarily prevent the table from functioning immediately, but can lead to complete failure of the table mounting mechanics. in this case, the table may drop resulting in potentially hazardous situations for patients and/or staff.
  • 조치
    Siemens is advising customers to avoid any collision of the table with objects in the table's operating area. Customers should contact their Siemens service representative for a precautionary inspection should such a collision occur or have occurred in the past.