IH-Com Kit Full version (Data management and Result Interpretation Software for ABO blood grouping). An in vitro diagnostic medical device (IVD)Used for ABO Blood GroupingAn in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Bio-Rad Laboratories Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-01215-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2012-11-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There have been reports of results not being interpreted in the anti-ab well (abo3) of id-cards intended for abo forward grouping, the ih-com software doesn't include the reaction of this well when sending the result to the laboratory host (result code-30).This can occur in the following situations;1. ih-com software is connected to the laboratory host through astm communication mode, and2. an "abo not interpretable" result is obtained due to a reaction "?" or "+-" or "dp" in the anti-ab well of the id-card intended for abo forward grouping, and3. the "second reading" option is deactivated and ih-com results are directly sent to the host, or4. the "second reading" option is activated by the the "not interpretable" result is obtained in ih-com directly saved and sent to host without a manual selection by the user in abo scrollbar.
  • 조치
    Bio-Rad Laboratories will be providing work around instructions in the customer letter. A software update will be implemented to correct the problem.

Device

  • 모델명 / 제조번호(시리얼번호)
    IH-Com Kit Full version (Data management and Result Interpretation Software for ABO blood grouping). An in vitro diagnostic medical device (IVD)Used for ABO Blood GroupingAn in vitro diagnostic medical device (IVD)
  • Manufacturer

Manufacturer