IMMULITE 2000 and IMMULITE 2000 XPi - GI-MA (CA 19-9) Positive Bias 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00131-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-02-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has confirmed a positive bias with the bio-rad lyphochek tumor marker plus control and the bio-rad liquichek tumor marker control when used with immulite 2000/immulite 2000 xpi gi-ma (ca 19-9) assay kit lots 312 and 313.The bias in quality control values is the result of the introduction of a new lot of the bead coat antibody (murine monoclonal anti-ca 19-9 antibody) in kit lot 312. when comparing kit lot 312 with kit lot 311, siemens observed an average percent bias of 38% (ranging from 30% to 47%) for patient samples recovering from 29.6 to 44.4 u/ml and an average percent bias of 5% (ranging from 1% to 11%) for patient samples recovering from 540 to 660 u/ml.Future kit lots are expected to recover quality control and patient samples similar to kit lot 312.
  • 조치
    Siemens is advising that patients may need to be re-base lined with kit lot 312 or higher. A look back over previously generated results should be discussed with the treating physician.

Device

  • 모델명 / 제조번호(시리얼번호)
    IMMULITE 2000 and IMMULITE 2000 XPi - GI-MA (CA 19-9) Positive BiasCatalogue number: L2KGI2Siemens Material Number: 10380988Lot numbers: 312 and 313ARTG number: 180780
  • Manufacturer

Manufacturer