Infinia, Infinia Hawkeye, Millenium VG, VG Hawkeye and VariCam Nuclear Medicine Systems, All configurations 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00143-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-02-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has become aware of a potential safety issue related to unintended radial detector motion, which may occur during patient setup or at the end of a patient scan. a potential life-threatening hazardous situation may occur if there is unintended radial detector motion leading to a patient contact. this failure mode has been observed on a single system for a whole body scan, in which detector moved to the home position and then slowly started drifting inward toward the patient. this generated both a visual and audio alert condition on both the console and the gantry.
  • 조치
    GE Healthcare is providing safety instructions to prevent any patient injuries if this failure mode occurs in the field. A GE Healthcare Service representative will perform the required service action (that includes a software update) on each affected system.

Device

  • 모델명 / 제조번호(시리얼번호)
    Infinia, Infinia Hawkeye, Millenium VG, VG Hawkeye and VariCam Nuclear Medicine Systems, All configurationsARTG Number: 200439
  • 의료기기 분류등급
  • Manufacturer

Manufacturer