Infinity Acute Care Monitoring System (IACS) used with Kontron C500 Cockpit Kontron C500 Cockpits with revision index 06 or higher that contain 4GB RAM Modules 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00735-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-06-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Dräger has learned that dräger infinity acute care systems that use the newer kontron cockpits with revision index 06 or higher that contain 4gb ram modules may not annunciate audio or visual alarms on the cockpit and central station due to high memory usage. the infinity m540 will still issue both audible and visual alarms; the alarm function is not affected at the infinity m540 and infinity central station when the cockpit is disconnected. to date, dräger are not aware of any negative patient consequences arising from this issue.
  • 조치
    Dräger will be contacting users to schedule a time to rework the identified devices by removing the affected Virtium RAM Modules and replacing them with the Swissbit version. In the interim, users are to remove the C500 Cockpit from service; the Infinity Acute Care System can be used without the C500 Cockpit as the Infinity M540 will function as the primary monitoring device once the Cockpit is disconnected.

Device

  • 모델명 / 제조번호(시리얼번호)
    Infinity Acute Care Monitoring System (IACS) used with Kontron C500 CockpitKontron C500 Cockpits with revision index 06 or higher that contain 4GB RAM ModulesMultiple Cockpit Serial NumbersARTG Number: 168054
  • 의료기기 분류등급
  • Manufacturer

Manufacturer