Infinity Acute Care Monitoring System (IACS) with software version VG5.0 and Microstream (Patient monitoring system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01412-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-11-01
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Draeger have become aware of two issues with the use of iacs software version vg5.0, as follows:1. it is possible that expected audible alarms may not occur at both the iacs cockpit and m540 although the appropriate audible and visual alarms will occur at the infinity central station (ics) and appropriate visual alarms will occur from the iacs system. this audible alarming anomaly could lead to missing expected audible alarms from iacs. the ics alarms (visual and audible) as intended and lacs will still display visual alarms.2. microstream mcable co2, which is supported by software version vg5.0, displays the incorrect unit of measure (uom) in contradiction to the labelling. the monitor indicates that the uom is mmhg, however, the actual value presented is in % volume. the incorrect uom could contribute to an error in the interpretation of the presented value. all distributed devices will have this anomaly.There have been no injuries reported for the above issues.
  • 조치
    Draeger will be contacting customers to arrange for downgrade of the IACS System to software version VG4. In the interim, customers can continue to use IACS software version VG5.0 but only if it is used in connection with an ICS or there is constant supervision of visual alarms at the bedside. The Microstream MCable is not to be used. This action has been closed-out on 28/06/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Infinity Acute Care Monitoring System (IACS) with software version VG5.0 and Microstream (Patient monitoring system)ARTG Number: 168054
  • Manufacturer

Manufacturer