Ingenuity CT, Ingenuity Core, Ingenuity Core128 with Continuous Computed Tomography (CCT) Option and have been upgraded from Software Version 3.x to Version 4.0.0 / 4.0.1 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01100-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-10-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips became aware through a customer complaint that while performing a cct examination on an ingenuity ct system, the first image was marked as “last shot” rather than the actual “last shot” acquired. after the clinician pressed the pedal for first exposure, images were displayed as the “last shot”. subsequent pedal presses did not result in immediate image reconstruction and the first image remained on the screen labelled as the “last shot”. eventually all images appeared. during an interventional procedure, there is a risk of serious injury to the patient if the clinician does not have real-time feedback of needle placement when the needle is near vital structures (major blood vessel, organ, etc).
  • 조치
    Philips is advising that affected users should check their CCT exam card and correct if the issue exists as described in the instructions supplied with the customer letter. A Philips Field Service Engineer will contact customers to schedule the software update at their site to permanently correct the issue. This action has been closed-out on 05/09/2016.