Innova 2121IQ and Innova 3131IQ Cardiovascular X-Ray Imaging Systems (diagnostic fluoroscopic angiography x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00802-3
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-08-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    While performing a fluoroscopic acquisition with the innova 2121iq or innova 3131iq systems, there is a potential for loss of usable imaging mode which requires the system to restart in order to recover. it may happen in a very specific scenario when one operator is requesting a lih (last image hold) functionality and another operator starts a short fluoroscopy exposure (stopping fluoroscopy exposure within 2 seconds) without waiting for the display of the lih (last image hold) in review monitor. the user would need to restart the system in order to regain acquisition capability.
  • 조치
    GE Healthcare is providing work around instructions to mitigate the risk and will be implementing a software update to permanently correct the issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    Innova 2121IQ and Innova 3131IQ Cardiovascular X-Ray Imaging Systems (diagnostic fluoroscopic angiography x-ray system)ARTG Number: 93871
  • Manufacturer

Manufacturer