Inspira Air Balloon Dilation System (Bronchial balloon catheter) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00907-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-09-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is potential for the balloon to not deflate or deflate slowly, which could result in airway obstruction until the balloon is deflated or removed. reports of difficulty deflating the balloon during a procedure have been received. in each of these reports, difficulty deflating occurred after the surgeon had pulled against resistance in response to the balloon moving distally during dilation. the force applied to the catheter stretched and narrowed the balloon catheter shaft, causing the balloon to be difficult or impossible to deflate. the potential for this issue to occur exists in all lots and all sizes of the inspira air balloon dilation system.
  • 조치
    Johnson & Johnson Medical is updating the information in the instructions for use (IFU) to include the following warnings: - Should the balloon move distally or proximally in the patient while inflated, it is not be held or pulled against. - The Inspira Air Balloon Catheter should only be placed in anatomical locations where a portion of the balloon can be continually visualised to ensure access in the unlikely event of deflation difficulty.

Device

  • 모델명 / 제조번호(시리얼번호)
    Inspira Air Balloon Dilation System (Bronchial balloon catheter)Product Codes: BC0524A, BC0524AZ, BC0724A, BC0724AZ, BC1040A, BC1040AZ, BC1440A and BC1440AZARTG Number: 172707
  • 의료기기 분류등급