Integrity R1.2 and Integrity R3.2 systems with a Precise Treatment Table for ATM license option 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Elekta Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01407-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-11-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Elekta have identified that it is possible to get a positional error with the precise treatment table following automatic table movement. this can occur if there is an undetected failure of the positional sensors. this has been previously communicated through notification 200-01-204-011 (rc-2014-rn-00927-1).It is possible to deliver treatment with the patient in an incorrect position, if this fault were to occur and go undetected.
  • 조치
    Recipients should distribute this notice to all users of the system and include a copy of the notice with the system manuals. The supplied Field Safety Notice serves to reiterate the importance of a procedure to check for the error already present in the IFU, previously communicated to customers (TGA Ref: RC-2014-RN-00927-1). Users should continue to follow the recommended daily QA checks detailed in the IFU, also present in the previous Field Safety Notice, until the software upgrade is released. Acknowledge the receipt of this notice by completing the Confirmation of receipt form you receive from Elekta's automated distribution system. Please return the form to Elekta even if you no longer have the affected units at your site. Elekta will release Integrity™ R4.0.0, which will identify positional errors over 5 mm if a sensor has failed. Upgrades in the field are expected to start in the first half of 2018. Elekta will contact users to plan and schedule this upgrade.

Device

  • 모델명 / 제조번호(시리얼번호)
    Integrity R1.2 and Integrity R3.2 systems with a Precise Treatment Table for ATM license optionARTG Number: 191025
  • Manufacturer

Manufacturer