IntelePACS(Image management system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00404-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-05-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Images successfully transmitted to intelepacs may not be available for reading in all studies, and the reader could be unaware of missing images.. this failure can be detected if the client has image count verification protocols. in addition if a large number of consecutive images are missing it will cause a noticeable gap in the anatomy.This issue can arises in two scenarios:1. for the three clients that employ ingest servers to provide temporary store-and-forward capabilities in their architecture, in some conditions, these servers fail to forward all received images to the permanent storage servers and no record of these images persist in the main archive. 2. for the remaining client architectures that do not employ temporary ingest servers, there is still a very low probability of failure for the device to properly replicate and register images when transferring them to another server.
  • 조치
    For the three clients using a configuration with ingest servers as temporary storage, an immediate update process was undertaken via the Client Specific Release process (completed as of 2015-04-23) to avoid the possibility of data loss. For the remaining clients, the normal Full Release Update (FRU) will be expedited. This action has been closed-out on 5/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    IntelePACS(Image management system)Software versions:4.7.1 R27 and up 4.8.1 R20 and up 4.9.1 R17 and up 4.10.1 R9 and up 4.11.1 R3 and up ARTG number: 149923
  • Manufacturer