IntelliSpace Portal with AutoSPECT Pro reconstruction application (image viewing system for diagnostic imaging, x-ray and computed tomography) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00364-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-03-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During spect reconstructions using attenuation correction and scatter correction, no scatter correction is being applied in the autospect pro application on intellispace portal with the software versions mentioned above. in addition, resolution recovery is not applied correctly in spect reconstructions using astonish in autospect pro on intellispace portal unless the data is acquired with vxgp collimators. the occurrence of either problem may result in image quality degradation that could result in misdiagnosis under certain conditions.
  • 조치
    Philips is providing work around instructions to ensure the continued safe use of the AutoSPECT Pro until a software update can be provided to correct the issue. This action has been closed-out on 12/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    IntelliSpace Portal with AutoSPECT Pro reconstruction application (image viewing system for diagnostic imaging, x-ray and computed tomography)Affected IntelliSpace Portal software versions: v5.0.0, v5.0.1, v5.0.2 and v6.0.0ARTG Number: 188684
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA