IntelliVue MX40 Software Versions: B.00.x and B.01.x 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00676-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-06-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The alarm pause time is user selectable from 1 to 3 minutes, with a default of 2 minutes, when operating the mx40 with the piic ix.When the following specific conditions are met, the alarm pause time does not automatically restart when it reaches its selected time limit:- the pause alarms configuration is enabled in the non-default setting of both red and yellow level alarms;- the mx40 is selected for the non-default setting of monitor mode;- the pause alarms is initiated to pause an active alarm or to maintain a short ‘non-alarm’ state while in the patient’s vicinity; and- an alarm condition occurs during the pause alarms time.The mx40 will continue to display “alarms paused: 0:00” and the piic ix will display “alarms paused” until the pause alarms feature is disabled manually.
  • 조치
    Philips is advising their customers that the following two items must be configured when using the MX40 with the IntelliVue Information Center iX: 1. Utilise the default state of Pause Alarms at Yellow only, which disables the ability to Pause/Suspend all Red level alarms. 2. Utilise the default state of Remote Suspend Alarms as off, which disables the use of the Pause Alarm feature of the MX40 device. Philips will implement a software update to correct the issue. This action has been closed-out on 12/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    IntelliVue MX40 Software Versions: B.00.x and B.01.xPart Numbers: 865350 and 865352ARTG Number: 99204