IntelliVue MX40 with software revisions B.05, B.06 and B.06.5X 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00494-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-05-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has identified that five warning statements are missing from the instructions for use (ifu) for intellivue mx40 ifu software revisions b.05, b.06 and b.06.5x. if users are unaware of the hazards or limitations described in the missing warning statements, they may not properly assess or provide appropriate treatment to a patient being monitored using an intellivue mx40.These ifus are missing warning statements related to monitoring of patients with pacemakers and the interpretation of qt/qtc measurements that were present in earlier revisions of the ifu. a “warning” alerts users to a potential serious outcome, adverse event or safety hazard. failure to observe a warning may result in death or serious injury to a user or patient.To date, no patient harm has been reported as a result of this issue.
  • 조치
    Phillips will contact affected customers to provide a copy of an Instructions For Use (IFU) Errata. Customers are requested to attach the Errata Sheet to the first page of Chapter 6 of the IFU.