Intuitive Harmonic ACE Instruments (used with the da Vinci Standard, S, and Si Surgical Systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00135-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-02-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Intuitive surgical has found that the harmonic ace ifu contains incorrect instructions in regard to the proper use of the instrument release kit (irk) (or “emergency grip release allen wrench”). the harmonic ace ifu instructs the user to turn the irk counter-clockwise. the correct instruction is to turn the irk clockwise.
  • 조치
    Users are asked to discard the current Harmonic Ace IFUs (IFU Part number: 550635-01, 550635-02 550635-03, 550635-04) that they may have. Users are asked to refer to either the Instruments and Accessories User Manual, the Quick Reference Guide that is physically attached to the system, and the metal instruction card that is attached to the IRK provide step-by-step instructions on how to properly perform the emergency grip release.Please also note that the Instruments and Accessories User Manual contain other warnings not listed in the IFU or have information listed in more detail. This action has been closed-out on 15/07/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Intuitive Harmonic ACE Instruments (used with the da Vinci Standard, S, and Si Surgical Systems)5mm Harmonic ACE instrumentsProduct Codes: INS-400274, INS-4002758mm Harmonic ACE instrumentsProduct Codes: INS-420274, INS-420275Harmonic ACE Curved Shears Disposable InsertProduct Code: INS-400272ARTG number: 132451